PRESCANBODY
Method that provides an initial reference thanks to a scan before/after proceeding with implant placement. The components used are: DAS PreScanbody and DAS Anchor Screw.


Phase preoperative preparation

NOTES
• Do not exceed the maximum torque of 20 N·cm to avoid bone damage or loss of system stability.
• If during the insertion of the component the required position is not reached without exceeding 20 N·cm torque, stop, carefully remove the screw, and reposition it again at 20 N·cm.
• It is recommended to have the surgical guide ready at this stage (in accordance with phase 3).
Phase preoperative scan

NOTES
• If the DAS PreScanbody interferes with the placement of the surgical guide, it can be removed and placed after the surgery. Due to its geometry, it will always retain its original position.
WARNING
Excessive torque beyond the recommended limits may fracture or damage the screw.